This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

COBAS 8000 DATA MANAGER (2015-05-22)

Starting date:
May 22, 2015
Posting date:
July 7, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-54010

Recalled Product

A. COBAS 8000 DATA MANAGER

Reason

Potential issue with the Cobas 8000 data manager system (all versions) where the option "automatic patient masking when QC error" option does work correctly. This functionality does not work if the results for the last measured control level for this test is valid. As a consequence, the results for the measured samples would not be released and forwarded to host due to the QC error.

Affected products

A. COBAS 8000 DATA MANAGER

Lot or serial number

N/A

Model or catalog number
  • 6888160001
  • 7267657001
Companies
Manufacturer

Roche Diagnostics Gmbh

Sandhoferstrasse 116

Mannheim

68305

GERMANY