This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Cobas 8000 - C502 Chemistry Analyzer (2014-03-27)

Starting date:
March 27, 2014
Posting date:
April 15, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-39089

Recalled products

  1. Cobas 8000 - C502 Chemistry Analyzer

Reason

The integrated cassette volume check (ICVC) is not automatically performed on the Cobas 80000-C502 Chemistry Analyzer for the Cobas C Pack Multi and for the Carbamazepine Carb3 for C311 and C501/502, due to ICVC being deactivated in the Rec file for these reagents.

Affected products

Cobas 8000 - C502 Chemistry Analyzer

Lot or serial number

Serial numbers:

  • 1019-08
  • 1019-10
  • 1128-05
  • 1140-04
  • 1247-08
Model or catalog number
  • 05964067001
Companies
Manufacturer

Roche Diagnostics GmbH

Sandhofer Strasse 116

Mannheim

68305

GERMANY