CMV IGM for ELECSYS and COBAS E and Diluent Universal (2015-02-18)
- Starting date:
- February 18, 2015
- Posting date:
- March 5, 2015
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type I
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-52463
Recalled Products
A. CMV IGM for ELECSYS and COBAS E
B. Diluent Universal
Reason
This recall is related to a potential reduction in the signal of the Elecsys CMV IGM assay when used in combination with some specific Lots of Diluent Universal which may lead to false negative results for patients samples with high reactivity against one specific antigen. This may be affected by reduced cut-off values which may lead to a false negative result when the affected combination is used. The signal reduction has only minor effects on the calibrators and controls.
Affected products
A. CMV IGM for ELECSYS and COBAS E
Lot or serial number
179689
182962
Model or catalog number
04784618190
Companies
- Manufacturer
-
Roche Diagnostics Gmbh
Sandhoferstrasse 116
Mannheim
68305
GERMANY
B. Diluent Universal
Lot or serial number
175913
176099
176899
178383
178386
180209
183652
183653
Model or catalog number
03183971122
11732277122
Companies
- Manufacturer
-
Roche Diagnostics Gmbh
Sandhoferstrasse 116
Mannheim
68305
GERMANY