Health product recall

CMV IGM for ELECSYS and COBAS E and Diluent Universal (2015-02-18)

Starting date:
February 18, 2015
Posting date:
March 5, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type I
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-52463

Recalled Products

A. CMV IGM for ELECSYS and COBAS E

B. Diluent Universal

Reason

This recall is related to a potential reduction in the signal of the Elecsys CMV IGM assay when used in combination with some specific Lots of Diluent Universal which may lead to false negative results for patients samples with high reactivity against one specific antigen. This may be affected by reduced cut-off values which may lead to a false negative result when the affected combination is used. The signal reduction has only minor effects on the calibrators and controls.

Affected products

A. CMV IGM for ELECSYS and COBAS E

Lot or serial number

179689

182962

Model or catalog number

04784618190

Companies
Manufacturer

Roche Diagnostics Gmbh

Sandhoferstrasse 116

Mannheim

68305

GERMANY



B. Diluent Universal

Lot or serial number

175913

176099

176899

178383

178386

180209

183652

183653

Model or catalog number

03183971122

11732277122

 

Companies
Manufacturer

Roche Diagnostics Gmbh

Sandhoferstrasse 116

Mannheim

68305

GERMANY