Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-67042

Affected products

A. Chemstrip 10A

B. Urisys 1100 Analyzer

Reason

It was discovered during internal performance studies that the Limit Of Detection (LOD) of the Chemstrip 10A needs to be changed since the current performance claim showed deviating values for protein, nitrite, ketone bodies, leukocytes, blood (intact erythrocytes) when measured on Urisys 1100 analyzer.