Centricity Universal Viewer - Main Unit (2018-02-06)
- Starting date:
- February 6, 2018
- Posting date:
- February 16, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-66010
Affected products
Centricity Universal Viewer - Main Unit
Reason
The following issue is being addressed by GE Healthcare File No. FMI 85442. GE Healthcare has recently become aware that a database handling error could occur during the image acquisition process affecting the completeness of acquired images with Centricity Universal Viewer running on a PACS-IW foundation. There is a potential that one or more "image series" (i.e. all images within an image set) may be missing from an exam without a user warning displayed in the Viewer. While this is rare, this can occur with imaging studies that consist of a very small number of images per series. E.g. CR Thorax exam with 1 image per series. There have been no injuries reported as a result of this issue.
Affected products
Centricity Universal Viewer - Main Unit
Lot or serial number
GTIN 00840682103800
Model or catalog number
2088034-X