Help us improve our website

Health product recall

Centricity Universal Viewer - Main Unit (2018-02-06)

Starting date:
February 6, 2018
Posting date:
February 16, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-66010

Affected products

Centricity Universal Viewer - Main Unit

Reason

The following issue is being addressed by GE Healthcare File No. FMI 85442. GE Healthcare has recently become aware that a database handling error could occur during the image acquisition process affecting the completeness of acquired images with Centricity Universal Viewer running on a PACS-IW foundation. There is a potential that one or more "image series" (i.e. all images within an image set) may be missing from an exam without a user warning displayed in the Viewer. While this is rare, this can occur with imaging studies that consist of a very small number of images per series. E.g. CR Thorax exam with 1 image per series. There have been no injuries reported as a result of this issue.

Affected products

Centricity Universal Viewer - Main Unit

Lot or serial number

GTIN 00840682103800

Model or catalog number

2088034-X

Companies
Manufacturer

GE Healthcare

540 West Northwest Hwy.

Barrington

60010

Illinois

UNITED STATES