This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

CELL-DYN EMERALD SYSTEM - CLEANER REAGENT (2016-03-29)

Starting date:
March 29, 2016
Posting date:
July 4, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-57846

Affected products

  1. CELL-DYN EMERALD SYSTEM - CLEANER REAGENT

Reason

They have identified occurrences where the CELL-DYN Emerald analyzer generates quality control (QC) out of range low for parameters RBC and PLT when using CELL-DYN Emerald cleaner lots# 6853, 6901, 6953, 6991, 7024, 7027, 7044, 7082, 7110 and 7119. Not all customers are experiencing this issue. Replacement of the CELL-DYN Emerald cleaner, along with a system cleaning of the instrument, must be performed in order to resolve the issue.

Affected products

  1. CELL-DYN EMERALD SYSTEM - CLEANER REAGENT

Lot or serial number
  • 6853
  • 6901
  • 6953
  • 6991
  • 7024
  • 7027
  • 7044
  • 7082
  • 7110
  • 7119
Model or catalog number
  • 09H46-02
Companies
Manufacturer

Abbott Laboratories Diagnostcis Division

100 Abbott Park Road,

Abbott Park

60064

Illinois

UNITED STATES