This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

CD8 PE (Leu 2a) Reagent (2014-12-22)

Starting date:
December 22, 2014
Posting date:
January 14, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-43229

Recalled Products

CD8 PE (Leu 2a) Reagent

Reason

May be hazardous to health: The CD8 PE reagent is contaminated with CD4 FITC. CD4 cells will be expressed along the FITC axis. Any marker that is stained for on the FITC axis will contain this incorrectly stained CD4 population. With subsequent down gating, the results will appear as if the CD4 population is not present. This could result in the unnecessary intensification of therapy for patients already known to have reductions in their CD4 levels.

Affected products

CD8 PE (Leu 2a) Reagent

Lot or serial number

4220622

Model or catalog number

340046

Companies
Manufacturer

Becton, Dickinson and Company

BD Biosciences

2350 Qume Drive

San Jose

95131

California

UNITED STATES