Catheter Extension
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
Catheter Extension |
More than 10 numbers, contact manufacturer. |
2843241 |
Issue
The manufacturer determined through testing that non-dioxin-like polychlorinated biphenyl acids (PCBAs) could leach from the stay•safe catheter extension, which are manufactured using peroxide cross-linked silicone tubing. Please note that PCBAs are chemically different from polychlorinated biphenyls (PCBs), which were not identified. To date, no complaints have been received related to PCBAs. PCBA levels decrease with shorter stay•safe catheter extension and decrease over the treatment time.
Recall start date: February 20, 2024
Additional information
Details
| Fresenius Medical Care Ag |
| Else-Kroner-Strasse 1, Bad Homburg, Germany, 61352 |
Get notified
Receive emails about new and updated recall and safety alerts.