This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Carescape R860 (Evair component) (2015-07-30)

Starting date:
July 30, 2015
Posting date:
August 11, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-54598

Affected Products

A. Carescape R860 (Evair component)

Reason

The following issue is being addressed by GE healthcare file no. FMI 34066: the EVair component manufacturer, Imtmedical AG, is conducting a field corrective action. Imtmedical has identified that potential friction of cables and tubing internal to the EVair compressor may result in an inability for the compressor to output compressed air. There is a potential for temporary oversupply of oxygen to the patient as a result of this failure because the GE Healthcare ventilators are designed to ventilate with 100% oxygen in the case of loss of input air supply pressure.

Affected products

A. Carescape R860 (Evair component)

Lot or serial number

>10, contact mfg.

Model or catalog number

1506-8600-000

Companies
Manufacturer
Datex-Ohmeda
Inc.
3030 Ohmeda Drive
Wisconsin
UNITED STATES