Health product recall

CareLink Software Suite (2020-10-28)



Starting date:
October 28, 2020
Posting date:
November 27, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-74407


Last updated:

Affected Products

A. Carelink Personal

B. Carelink Software

Reason

Medtronic released an update to the CareLink software suite on Saturday, August 29, 2020. Following this release, customers who generated a Device Settings report from CareLink Personal or CareLink system software may have been presented with an inaccurate carbohydrate ratio setting on their CareLink report. This only occurred under the following circumstances where:

  1. The carbohydrate unit of measurement in CareLink Software is set to grams, but the insulin pump is set to exchanges.
  2. The carbohydrate unit of measurement in CareLink Software is set to exchanges, but the insulin pump is set to grams.

If adjustments were made to the carbohydrate ratio based on the CareLink Device Settings report, this could lead to either over or under delivery of insulin. This could cause low blood glucose (hypoglycemia, loss of consciousness, seizure, death), and/or high blood glucose (hyperglycemia).

On September 30, 2020, Medtronic released an automatic update to CareLink software which resolved the issue described above. However, some customers may have adjusted their carbohydrate ratio between August 29, 2020 and September 30, 2020 using incorrect data on the Device Settings report.

Affected products

A. Carelink Personal

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

MMT-7333WW

Companies
Manufacturer

Medtronic Minimed

18000 Devonshire St.

Northridge

91325

California

UNITED STATES



B. Carelink Software

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

MMT-7350

Companies
Manufacturer

Medtronic Minimed

18000 Devonshire St.

Northridge

91325

California

UNITED STATES