Health product recall

CareLink SmartSync™ Applications

Brand(s)
Last updated

Summary

Product
CareLink SmartSync™ Applications
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

CareLink SmartSync™ Device Manager Application

All lots.

CARELINK SMARTSYNC DEVICE MANAGER APP

CareLink SmartSync™ Azure™ Astra™ Application

All lots.

D00U003

CareLink SmartSync™ Cobalt™ Crome™ Application

All lots.

D00U005

CareLink SmartSync™ Evera™ MRI Application

All lots.

D00U010

CareLink SmartSync™ Micra™ AV Application

All lots.

D00U007

CareLink SmartSync™ Viva™ Consulta™ Syncra™ Advisa™ Ensura™ Application

All lots.

D00U008

CareLink SmartSync™ PSA Application

All lots.

D00U002

CareLink SmartSync™ VisiaAF™ Application

All lots.

D00U011

CareLink SmartSync™ Viva™ Brava™ Evera™ Application

All lots.

D00U012

CareLink SmartSync™ Micra™ VR Application

All lots.

D00U006

CareLink SmartSync™ Micra™ VR2 Application

All lots.

D00U022

CareLink SmartSync™ Micra™ VR2 Application

All lots.

D00U009

CareLink SmartSync™ Micra™ AV2 Application

All lots.

D00U022

CareLink SmartSync™ Percepta™ Serena™ Solara™ Application

All lots.

D00U004

Issue

This recall pertains to a device electrical reset message being displayed inappropriately (i.e., when a true device electrical reset did not occur). There are two issues when the device electrical reset message is displayed inappropriately:

  • Issue #1: The user is informed there was a device electrical reset when there was not. This can cause confusion and might lead the user to believe the device is no longer functioning properly. 
  • Issue #2: One of the messages that can be displayed inappropriately recommends device explant ("serious device error: device should be explanted"). While the message may be factually correct for the type of electrical reset it is designed to indicate, if displayed inappropriately it increases the risk of explant and patient harm.

Recall start date: June 18, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Health products - Medical devices - Cardiovascular
Companies

Medtronic Inc.

710 Medtronic Parkway N.E., Minneapolis, Minnesota, United States, 55432

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-77646

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe