Health product recall

CARELINK SMARTSYNC DEVICE MANAGER (2020-12-18)



Starting date:
December 18, 2020
Posting date:
January 7, 2021
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-74709


Last updated:

Affected Products 

CARELINK SMARTSYNC DEVICE MANAGER

Reason

This update addresses a rare communication sequence during the first device interrogation with a Smartsync device manager that may result in the temporary suspension of some device features (i.e., battery measurements, capture management, atrial lead position check, effectivcrt algorithms, and adaptivcrt).This rare interaction results in temporary suspension of automatic threshold testing and output adjustments, and suspension of auto-optimization of CRT therapy. The issue is unlikely to result in clinical impact to the patient, and features are restored upon next programmer device interrogation or presence of a magnet.

Affected products

CARELINK SMARTSYNC DEVICE MANAGER

Lot or serial number

All lots.

Model or catalog number

24970A

Companies
Manufacturer

Medtronic Inc.

710 Medtronic Parkway

Minneapolis

55432

Minnesota

UNITED STATES