Cardiosave System - Battery Only (2021-09-23)
- Starting date:
- September 23, 2021
- Posting date:
- October 1, 2021
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-76565
Last updated:
2021-10-01
Affected Products
Cardiosave System - Battery Only
Reason
A cardiosave Lithium-ION battery pack (0146-00-0097) used during protocol testing failed to meet the minimum runtime requirement per getinge's internal product specification. These nonconforming batteries were inadvertently released to customers.
Affected products
Cardiosave System - Battery Only
Lot or serial number
181725430IP
Model or catalog number
0998-00-0800-XX
Companies
- Manufacturer
-
Datascope Corp.
1300 Macarthur Blvd.
Mahwah
07430
New Jersey
UNITED STATES