CardioChek System - PTS Panels CHOL+HDL+GLU test strips (2019-04-11)
- Starting date:
- April 11, 2019
- Posting date:
- June 7, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-70101
Last updated:
Affected Products
CardioChek System - PTS Panels CHOL+HDL+GLU test strips
Reason
It has been determined that a triple-fault condition may cause a less-snug fit, or looseness of the strip within the guides on the analyzer. If the test strip "slips" from its fully inserted position, it may lead to the test not completing or an error code.
Affected products
CardioChek System - PTS Panels CHOL+HDL+GLU test strips
Lot or serial number
I811
I812
I814
Model or catalog number
2412
Companies
- Manufacturer
-
Polymer Technology Systems, Inc.
7736 Zionsville Rd
Indianapolis
46268
Indiana
UNITED STATES