Health product recall

CardioChek System - PTS Panels (Cholesterol Test Strips, CHOL+HDL+GLU Test Strips) (2019-03-26)

Starting date:
March 26, 2019
Posting date:
April 18, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-69624


Last updated:

Affected Products

  1. CardioChek System - PTS Panels Cholesterol Test Strips
  2. CardioChek System - PTS Panels CHOL+HDL+GLU Test Strips

Reason

It has been determined that a triple-fault condition may cause a less-snug fit, or looseness of the strip within the guides on the analyzer. In extreme cases, the test strip could "slip" from its fully inserted position. Investigation has shown that the issue occurs only as a triple-fault condition.

  1. Older analyzers that are still in use in the filed, may have additional leeway within the strip guides, due to extensive use. This is a normal part of aging.
  2. Strip carriers (bases that are used in test strip assembly) that trend to the lower end of the specification for thickness (but are still within specification), and
  3. The analyzer is not kept on a flat, stable surface during testing, or is physically moved during testing.

Affected products

A. CardioChek System - PTS Panels Cholesterol Test Strips

Lot or serial number

P832

Model or catalog number

1710

Companies
Manufacturer

Polymer Technology Systems, Inc.

7736 Zionsville Rd

Indianapolis

46268

Indiana

UNITED STATES



B. CardioChek System - PTS Panels CHOL+HDL+GLU Test Strips

Lot or serial number
  • I808
  • I811
  • I812
  • I814
Model or catalog number

2412

Companies
Manufacturer

Polymer Technology Systems, Inc.

7736 Zionsville Rd

Indianapolis

46268

Indiana

UNITED STATES