This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Capture-R Ready Indicator Red Cells

Starting date:
August 27, 2012
Posting date:
September 17, 2012
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public (GP), Healthcare Professionals, Hospitals
Identification number:
RA-15542

Recalled Products

A. Capture-R Ready Indicator Red Cells

Reason

Post-Release testing has demonstrated that the affected Lot (221861) may result in higher rates for equivocal and unexpected positive reactions compared to other in-date or previously released Lots.

Affected products

A. Capture-R Ready Indicator Red Cells

Lot or serial number

221861

Model or catalog number

0006427, '0006428

Companies
Manufacturer
Immucor Inc.