This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Capture-R Ready-ID (2016-05-25)

Starting date:
May 25, 2016
Posting date:
May 31, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-58702

Affected products

A. Capture-R Ready-ID

Reason

Immucor has received reports that plates of Capture-R Ready-Id (lot id 296) are being aborted due to "measurement out of range" errors after the strip clean read on the Galileo NEO instrument, due to cell 13 exhibiting a yellow color.

Affected products

A. Capture-R Ready-ID

Lot or serial number

Lot: ID296/Exp:2016-05-31

Model or catalog number

66214

Companies
Manufacturer

Immucor Inc.

3130 Gateway Drive, P.O. Box 5625

Norcross

30091-5625

Georgia

UNITED STATES