Capture-R Ready-ID (2016-05-25)
- Starting date:
- May 25, 2016
- Posting date:
- May 31, 2016
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-58702
Affected products
A. Capture-R Ready-ID
Reason
Immucor has received reports that plates of Capture-R Ready-Id (lot id 296) are being aborted due to "measurement out of range" errors after the strip clean read on the Galileo NEO instrument, due to cell 13 exhibiting a yellow color.
Affected products
A. Capture-R Ready-ID
Lot or serial number
Lot: ID296/Exp:2016-05-31
Model or catalog number
66214
Companies
- Manufacturer
-
Immucor Inc.
3130 Gateway Drive, P.O. Box 5625
Norcross
30091-5625
Georgia
UNITED STATES