Camino Advanced Monitor (2013-11-27)
- Starting date:
- November 27, 2013
- Posting date:
- January 6, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-37281
Recalled products
- Camino Advanced Monitor
Reason
Integra Lifesciences has identified that if the optional SYS/DIAS mode is used on Camino advanced monitors (CAM01) with serial numbers formatted as MExxxxx04, the displayed values for the systolic and diastolic intracranial pressure (ICP) are reversed. But, if the CAM01 is used with an auxiliary bedside monitor, the bedside monitor displays these values correctly.
Affected products
A. Camino Advanced Monitor
Lot or serial number
- MEA1301804
- MEC1300404
- MED1301104
- MED1301204
- MED1301304
- MEE1300504
- MEE1301504
Model or catalog number
- CAM01
Companies
- Manufacturer
-
Integra Neurosciences Ltd.
also trading as Spembly Medical
Newbury Road
Andover, Hampshire
SP10 4DR
UNITED KINGDOM