CADD Medication Cassette Reservoir,50ml (2015-01-28)
- Starting date:
- January 28, 2015
- Posting date:
- February 16, 2015
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-43735
Recalled Products
A) CADD Medication Cassette Reservoir,50ml
Reason
Smiths Medical has received a small number of complaints of leakage in some CADD medication cassette reservoirs, 50ml. A small amount of leaking has been found to occur at the welding of the tube to the internal reservoir bag on some cassettes. If noticed during cassette filling, it could result in a delay of therapy while an alternative cassette is obtained. If noticed during fluid delivery with the patient, it could result in an interruption in therapy while the cassette is replaced. Smiths Medical has received no reports of serious injury or death related to this issue.
Affected products
A) CADD Medication Cassette Reservoir,50ml
Lot or serial number
14X-297
14X-323
14X-324
Model or catalog number
21-7001-24
21-7301-24
Companies
- Manufacturer
-
Smiths Medical ASD, Inc.
1265 Grey Fox Road
St. Paul
55112
Minnesota
UNITED STATES