Health product recall

CADD Medication Cassette Reservoir,50ml (2015-01-28)

Starting date:
January 28, 2015
Posting date:
February 16, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-43735

Recalled Products

A) CADD Medication Cassette Reservoir,50ml

Reason

Smiths Medical has received a small number of complaints of leakage in some CADD medication cassette reservoirs, 50ml. A small amount of leaking has been found to occur at the welding of the tube to the internal reservoir bag on some cassettes. If noticed during cassette filling, it could result in a delay of therapy while an alternative cassette is obtained. If noticed during fluid delivery with the patient, it could result in an interruption in therapy while the cassette is replaced. Smiths Medical has received no reports of serious injury or death related to this issue.

Affected products

A) CADD Medication Cassette Reservoir,50ml

Lot or serial number

14X-297

14X-323

14X-324

Model or catalog number

21-7001-24

21-7301-24

Companies
Manufacturer

Smiths Medical ASD, Inc.

1265 Grey Fox Road

St. Paul

55112

Minnesota

UNITED STATES