Health product recall

CADD Blue-Striped Administration Set (2018-02-28)

Starting date:
February 28, 2018
Posting date:
March 23, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-66270

Affected products

CADD Blue-Striped Administration Set

Reason

A detailed investigation points to a unique use case, involving a single Canadian customer using non-approved Smiths Medical pouches, with the CADD Blue-Striped Administration Set. Use of non-approved accessories, as indicated in labeling, may adversely affect pump operation. Smiths Medical has decided to proceed with sending communication to notify all Canadian customers who have purchased the CADD Blue-Striped Administration Set. The content of the Field Safety Notice (FSN) provides more information to the customer/user when using the administration set. FSN emphasizes the correct use of the product and how to avoid strain relief which in turn prevents leakage.

Affected products

CADD Blue-Striped Administration Set

Lot or serial number

More than 1000 numbers, contact manufacturer.

Model or catalog number

21-7036-01

Companies
Manufacturer
Smiths Medical ASD, Inc.
1265 Grey Fox Road
St. Paul
55112
Minnesota
UNITED STATES