This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

C-SERIES CARRY BAR

Starting date:
September 16, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-60452

Affected Products

Reason

C-Series Carry Bar have separated from the lift strap due to the metal strap pin breaking the black plastic puck and pulling through the top plate. This breaking has been linked to users not performing pre-use, monthly & bi-annual inspections of their lifting equipment as specified in all prism medical user manuals.  The black plastic puck does not support the weight of the client, but instead acts as a swivel and washer that is sandwiched between the metal carry bar top plate and metal strap pin. They are not aware of any defects that will/may cause the black plastic puck to break unexpectedly.

Affected products

C-SERIES CARRY BAR

Lot or serial number

Not Applicable

Model or catalog number
  • 360741
  • 360750
  • 360751
  • 360755
  • 360755 REV B
  • 360756
  • 360757
Companies
Manufacturer
Prism Medical USA
10888 Metro Ct.
Maryland Heights
63043
Missouri
UNITED STATES