This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Bubble CPAP System, Components and Nasal Prong (2014-04-11)

Starting date:
April 11, 2014
Posting date:
May 5, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-39317

Recalled Products

A. Bubble CPAP System

B. Bubble CPAP System - Components

C. Nasal Prong

Reason

FPH (Fisher and Paykel Healthcare) introduced a new version of Nasal CPAP prongs in 2013. Since the release of this new version, FPH has received reports of the affected prongs detaching from the nasal tubing use, especially in the presence of mucous and/or moisture. The nasal tubing is not affected by this recall.

Affected products

A. Bubble CPAP System

 

Lot or serial number

131007-140325

 

Model or catalog number

BC461-SK

BC471-SK

 

Companies
Manufacturer

Fisher and Paykel Healthcare Ltd.

15 Maurice Paykel Place East Tamaki, P.O. Box 14348 Panmure

Auckland

2013

NEW ZEALAND



B. Bubble CPAP System - Components

 

Lot or serial number

131007-140325

 

Model or catalog number

BC490-SK

BC491-SK

BC492-SK

 

Companies
Manufacturer

Fisher and Paykel Healthcare Ltd.

15 Maurice Paykel Place East Tamaki, P.O. Box 14348 Panmure

Auckland

2013

NEW ZEALAND



C. Nasal Prong

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

More than 10 numbers, contact manufacturer

Companies
Manufacturer

Fisher and Paykel Healthcare Ltd.

15 Maurice Paykel Place East Tamaki, P.O. Box 14348 Panmure

Auckland

2013

NEW ZEALAND