Health product recall

Brilliance ICT Systems and Ingenuity CT System (2016-10-12)

Starting date:
September 12, 2016
Posting date:
November 3, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-60804

Affected Products

  1. Brillance ICT System
  2. Brillance ICT SP system-Main
  3. Ingenuity CT system

Reason

The Philips-Granite product label, which is applied by the supplier (Granite) to the rear side of the IMPR (single computer) host cabinet (4598-004-44981), has incorrect current rating label applied. The granite label on the host IMR rack should have electrical rating values of: 230V-2.4 A, 50/60 Hz. The non-conforming label states the current rating as 4.8 A (instead of 2.4 A) , as defined in the host rack IMR specifications.

Affected products

A. Brillance ICT System

Lot or serial number
  • 100638
Model or catalog number
  • 728306
Companies
Manufacturer
PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC.
595 MINER ROAD
CLEVELAND
44143
Ohio
UNITED STATES

B. Brillance ICT SP system-Main

Lot or serial number
  • 200208
  • 200213
  • 30003
  • 30004
  • 320227
Model or catalog number
  • 728311
Companies
Manufacturer
PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC.
595 MINER ROAD
CLEVELAND
44143
Ohio
UNITED STATES

C. Ingenuity CT system

Lot or serial number
  • 30003
  • 30004
  • 320227
Model or catalog number
  • 728323
  • 728326
Companies
Manufacturer
PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC.
595 MINER ROAD
CLEVELAND
44143
Ohio
UNITED STATES