This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Brilliance ICT Systems and Ingenuity CT System (2016-10-12)

Starting date:
September 12, 2016
Posting date:
November 3, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-60804

Affected Products

  1. Brillance ICT System
  2. Brillance ICT SP system-Main
  3. Ingenuity CT system

Reason

The Philips-Granite product label, which is applied by the supplier (Granite) to the rear side of the IMPR (single computer) host cabinet (4598-004-44981), has incorrect current rating label applied. The granite label on the host IMR rack should have electrical rating values of: 230V-2.4 A, 50/60 Hz. The non-conforming label states the current rating as 4.8 A (instead of 2.4 A) , as defined in the host rack IMR specifications.

Affected products

A. Brillance ICT System

Lot or serial number
  • 100638
Model or catalog number
  • 728306
Companies
Manufacturer

PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC.

595 MINER ROAD

CLEVELAND

44143

Ohio

UNITED STATES



B. Brillance ICT SP system-Main

Lot or serial number
  • 200208
  • 200213
  • 30003
  • 30004
  • 320227
Model or catalog number
  • 728311
Companies
Manufacturer

PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC.

595 MINER ROAD

CLEVELAND

44143

Ohio

UNITED STATES



C. Ingenuity CT system

Lot or serial number
  • 30003
  • 30004
  • 320227
Model or catalog number
  • 728323
  • 728326
Companies
Manufacturer

PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC.

595 MINER ROAD

CLEVELAND

44143

Ohio

UNITED STATES