Health product recall

Brilliance Ict Sp System - Main

Starting date:
June 30, 2017
Posting date:
August 14, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-64120

Reason

  • Issue 1: Scan parameters exceeds system resources: During a bolus tracking procedure: If the operator sets the ROI on the locator slice and then presses "go" to initiate the tracker scans, the following message may display: "Scan parameters exceeds system resources". If the message is displayed, the system will fail to proceed with the tracker scan.
  • Issue 2: Scan may start at the top of the head, which may not be where the operator is expecting the scan to start: If the operator plans to do multiple scans with multiple results including identical links, then the first result of the subsequent acquisition will follow the last result of previous acquisition and not the identical link of the previous acquisition. The possible outcome is that the results of the second scan could be planned outside the boundaries of the expected results.
  • Issue 3: Error message appears stating: "The exam application stopped working" and will instruct the operator to close the program: During a CCT procedure an error message may appear "The exam application stopped working, close the program". When the operator closes the program a 'System Error' will appear and the application will fail and require restart.
  • Issue 4: During a bolus tracking procedure, after acquiring the surview, locator and tracker, the scanner may move into place, give the patient breathing instructions, and then fail to acquire the scan: During a bolus tracking procedure with a clinical scan that exceeds 118 seconds in duration, the following may occur: Once the bolus tracking threshold is reached, the couch will move into the planned positon for the clinical scan, but fail to acquire the scan.
  • Issue 5: Image volumes are not found in patient directory when thin images are included in one of the multiple results: the operator completes a CT orbit procedure and clicks 'End Study'. The final thin images may not appear in the patient directory. When the operator clicks 'recon' to attempt to do an off-line recon, the 'view 2' application may fail.
  • Issue 6: Application may fail when modifying the results plan box during 3D calcium score: If the operator attempts to modify the plan box of a result on the surview, while setting up a 3D calcium score scan, the application may fail and require a restart.
  • Issue 7: Tube too hot: The operator presses 'Go' to execute a timed scan sequence and the following message may display: "Tube too hot".
  • Issue 8: Memory overflow: When performing a CT clinical procedure, it is possible that the scan may be aborted as a result of a memory overflow issue.
  • Issue 9: While the reconstruction is in progress disconnecting the recon box (CIRS) may not give error message: When the connection between host computer and CIRS is disrupted during reconstruction, the operator may not be informed. This may require that the operator reconstruct the images through off-line reconstruction.
  • Issue 10: In the plan viewer if the start position is modified the total length of the acquisition will change as well: In the plan viewer, if the start position is modified by the length of one slice thickness, then the total length of the acquisition may change by the same length value. Note this occurs only when planning without a surview.

Affected products

Brilliance Ict Sp System - Main

Lot or serial number

  • 100101
  • 100195
  • 100223
  • 100440
  • 100616
  • 100619
  • 100638
  • 200208
  • 200213

Model or catalog number

728311

Companies

Manufacturer

Philips Medical Systems (Cleveland) Inc.

595 MINER ROAD

CLEVELAND

44143

Ohio

UNITED STATES