This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Brilliance CT Systems and Ingenuity CT System (2014-04-01)

Starting date:
April 1, 2014
Posting date:
April 15, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-39103

Recalled products

  1. Brilliance 40/64 Slice CT System
  2. Brilliance CT Big Bore System - Main Unit
  3. Brilliance 64 Slice CT System - Main Unit
  4. Brilliance ICT System - System
  5. Brilliance 6/10/16 Slice CT System
  6. Brilliance CT System (16P)
  7. Brilliance ICT SP System - Main Unit
  8. Ingenuity CT System

Reason

Philips has become aware that if the service latch is not properly secured during servicing of the device, the table top's subframe becomes free floating causing unintended, horizontal motion. To date, there have been no reported injuries or deaths resulting from this issue.

Affected products

A. Brilliance 40/64 Slice CT System

Lot or serial number
  • 9164
  • 9186
  • 9200
  • 9203
  • 9207
Model or catalog number
  • 4550 110 04011
Companies
Manufacturer

Philips Medical Systems (Cleveland), Inc.

595 Miner Road

Cleveland

44143

Ohio

UNITED STATES



B. Brilliance CT Big Bore System - Main Unit

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • 4535 670 88051
Companies
Manufacturer

Philips Medical Systems (Cleveland), Inc.

595 Miner Road

Cleveland

44143

Ohio

UNITED STATES



C. Brilliance 64 Slice CT System - Main Unit

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • 4550 110 00921
Companies
Manufacturer

Philips Medical Systems (Cleveland), Inc.

595 Miner Road

Cleveland

44143

Ohio

UNITED STATES



D. Brilliance ICT System - System

Lot or serial number
  • 100046
  • 100101
  • 100195
  • 100206
  • 100223
  • 100249
  • 100290
  • 100440
Model or catalog number
  • 728306
Companies
Manufacturer

Philips Medical Systems (Cleveland), Inc.

595 Miner Road

Cleveland

44143

Ohio

UNITED STATES



E. Brilliance 6/10/16 Slice CT System

Lot or serial number
  • 4010
  • 5074
  • 5092
  • 5141
  • 5151
  • 5158
  • 6352
Model or catalog number
  • 4535 670 73181
  • 4535 670 78851
Companies
Manufacturer

Philips Medical Systems (Cleveland), Inc.

595 Miner Road

Cleveland

44143

Ohio

UNITED STATES



F. Brilliance CT System (16P)

Lot or serial number
  • 6024
  • 6072
  • 6103
  • 6149
  • 6150
Model or catalog number
  • 453567005721
Companies
Manufacturer

Philips Medical Systems (Cleveland), Inc.

595 Miner Road

Cleveland

44143

Ohio

UNITED STATES



G. Brilliance ICT SP System - Main Unit

Lot or serial number
  • 200071
Model or catalog number
  • 728311
Companies
Manufacturer

Philips Medical Systems (Cleveland), Inc.

595 Miner Road

Cleveland

44143

Ohio

UNITED STATES



H. Ingenuity CT System

Lot or serial number
  • 320123
  • 320148
Model or catalog number
  • 728323
Companies
Manufacturer

Philips Medical Systems (Cleveland), Inc.

595 Miner Road

Cleveland

44143

Ohio

UNITED STATES