Brilliance CT Big Bore System-Main Unit (2018-01-05)
- Starting date:
- January 5, 2018
- Posting date:
- January 29, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-65736
Affected products
Brilliance CT Big Bore System-Main Unit
Reason
Philips became aware of issues with Brilliance Big Bore CT systems running software version 4.2.0. The most significant issues are:
- Issue #1: EFOV (extended field of view) check message does not display when FOV is set > 600mm for a paused scan
- Issue #2: Prefer “center x/y 0” and “disable result rotation” checkboxes checked automatically when checking the ""CT Simulation""
- Issue #3: Tumor LOC allows POI (Point of Interest) locking when a non-isocenter slice is displayed
- Issue #4: Tracker scan completed, no images were generated
- Issue #5: Continuous Link plans the clinical scan outside the Surview
- Issue #6: Applying tilt on first mpr result, when the other axial results are unplanned causes change in scan length and number of cycles of scan
- Issue #7: “System is still initializing. Please wait…” error
- Issue #8: Auto locations not functioning as expected
- Issue #9: System ended exam before completion
- Issue #10: System Freeze-Up after Pause is pressed
Affected products
Brilliance CT Big Bore System-Main Unit
Lot or serial number
Not applicable
Model or catalog number
728244
Companies
- Manufacturer
-
Philips Medical Systems (Cleveland) Inc.
595 Miner Road
Cleveland
44143
Ohio
UNITED STATES