This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

BRILLIANCE AND INGENUITY CT SYSTEMS (2015-07-30)

Starting date:
July 30, 2015
Posting date:
August 18, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-54688

Affected Products

  1. BRILLIANCE 40/64 SLICE CT SYSTEM
  2. INGENUITY CT SYSTEM

Reason

Some scans unexpectedly set the axial (Z axis) scan length to a greater length than the user set in the plan. The issue occurs only when the following 3 conditions are met:-the plan employs dual-surview, the plan requests either a sagittal or a coronal reconstruction, the gantry is tilted

Affected products

A. BRILLIANCE 40/64 SLICE CT SYSTEM

Lot or serial number
  • 9920
  • 9923
Model or catalog number

4550 110 04011

Companies
Manufacturer

Philips Medical Systems (Cleveland) Inc.

595 MINER ROAD

CLEVELAND

Ohio

UNITED STATES



B. INGENUITY CT SYSTEM

 

Lot or serial number

30003

30004-1

320123

32021

320227

Model or catalog number

728323

728326

Companies
Manufacturer

Philips Medical Systems (Cleveland) Inc.

595 MINER ROAD

CLEVELAND

Ohio

UNITED STATES