This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Brilliance and Ingenuity CT and ICT System (2016-04-21)

Starting date:
April 21, 2016
Posting date:
May 17, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
Healthcare Professionals, General Public, Hospitals
Identification number:
RA-58356

Affected products 

  1.  BRILLIANCE 40/64 SLICE CT SYSTEM

  2. BRILLIANCE ICT SYSTEM - SYSTEM

  3. BRILLIANCE ICT SP SYSTEM - MAIN

  4. INGENUITY CT SYSTEM

Reason

Philips has become aware of the following seven software issues:  1.Sagittal result shortened for axial scans, 2.Single series displayed in console viewer when creating MPR's., 3. Scan length changes with a change in field of view, 4. Different CTDI values with dose right, 5. Scan length changes on subsequent axial results, 6. Surview scan lengths near 135mm or 184mm may result in unexpected scan start positions, and 7. Reconstruction length does not equal scan length.

Affected products

A. BRILLIANCE 40/64 SLICE CT SYSTEM

 

Lot or serial number

All lots

Model or catalog number
  • 4550 110 09021
Companies
Manufacturer

Philips Medical Systems (Cleveland) Inc.

595 MINER ROAD

CLEVELAND

Ohio

UNITED STATES



B. BRILLIANCE ICT SYSTEM - SYSTEM

 

Lot or serial number

All lots

Model or catalog number
  • 728306
Companies
Manufacturer

Philips Medical Systems (Cleveland) Inc.

595 MINER ROAD

CLEVELAND

Ohio

UNITED STATES



C. BRILLIANCE ICT SP SYSTEM - MAIN

 

Lot or serial number

All lots

Model or catalog number
  • 728311
Companies
Manufacturer

Philips Medical Systems (Cleveland) Inc.

595 MINER ROAD

CLEVELAND

Ohio

UNITED STATES



D. INGENUITY CT SYSTEM

Lot or serial number

All lots

Model or catalog number
  • 728323
  • 728326
Companies
Manufacturer

Philips Medical Systems (Cleveland) Inc.

595 MINER ROAD

CLEVELAND

Ohio

UNITED STATES