This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Brilliance 40/64 Slice CT System, Ingenuity CT System (2016-08-29)

Starting date:
August 29, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-60148

Affected Products

  1. Brilliance 40/64 Slice CT System
  2. Ingenuity CT System

Reason

Philips has become aware that while performing high resolution head scans, the reconstructed images may exhibit degraded image quality manifested as:

  • non uniformity of the image
  • CT number shift, greater than 5 Hounsfield units
  • reduction in gray/white matter differentiation

the issue is most severe in the high resolution mode. Philips provides factory reference pediatric scan protocols which are set to high resolution by default. Adult head scans have a factory reference protocol that defaults to standard resolution which does not experience the issue

Affected products

A. Brilliance 40/64 Slice CT System

Lot or serial number
  • 10213
  • 9074
  • 9920
  • 9923
Model or catalog number
  • 4550 110 09021
Companies
Manufacturer

Philips Medical Systems (Cleveland) Inc.

595 MINER ROAD

CLEVELAND

Ohio

UNITED STATES



B. Ingenuity CT System

Lot or serial number

30003

30004

320123

32021

320227

Model or catalog number

728323

728326

Companies
Manufacturer

Philips Medical Systems (Cleveland) Inc.

595 MINER ROAD

CLEVELAND

Ohio

UNITED STATES