This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Brilliance 40/40 Slice CT System and Ingenuity CT System (2017-10-12)

Starting date:
October 12, 2017
Posting date:
October 27, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Hospitals, Healthcare Professionals
Identification number:
RA-64960

Affected Products

  1. BRILLIANCE 40/40 SLICE CT SYSTEM
  2. INGENUITY CT SYSTEM

Reason

Issue 1: Cardiac Signal not found in Workflow Issue-

If the operator loads a Cardiac Exam Card, the cardiac signal is shown during cardiac scan but not shown during the surview.

Issue 2: Scan may start at the top of the head, which may not be where the operator is expecting the scan to start. If the operator plans to do multiple scans with multiple results including identical links, then the first result of the subsequent acquisition will follow the last result of previous acquisition and not the identical link of the previous acquisition. The possible outcome is that the results of the second scan could be planned outside the boundaries of the expected results.

Issue 3: Pulmo Gates on Cardiac Workflow Issue -

If the operator loads a Cardiac Exam Card after performing a Pulmonary scan, the Pulmo wave may be present.

Issue 4: In the Plan viewer if the start position is modified the total length of the acquisition will changes as well. In the Plan viewer if the start position is modified by the length of one slice thickness, then the total length of the acquisition may change by the same length value as well. Note this occurs only when planning without a surview.

Issue 5: Image volumes are not found in patient directory when thin images are included in one of the multiple results. The operator completes a CT Orbit procedure and clicks ‘End Study’. The final thin images may not appear in Patient Directory. When the operator clicks ‘Recon’ to attempt to do an off-line recon the ‘View 2’ application may fail.

Issue 6: Application may fail when modifying the results Plan Box during 3D Calcium Score

While setting up a 3D Calcium Score scan, if the Operator attempts to modify the plan box of a result on the surview, the application may fail and require a restart.

Issue 7: Tube too hot-

The Operator presses ‘Go’ to execute a timed scan sequence and the following message may display: “Tube Too Hot”.

Issue 8: Memory overflow Issue-

When performing a CT clinical procedure, it is possible that the scan may be aborted as a result of a memory overflow issue.

Issue 9: While the reconstruction is in progress disconnecting the Recon box (CIRS) may not give error message. When the connection between Host computer and recon box (CIRS) is disrupted during reconstruction, the operator may not be informed and may require that the operator reconstruct the images through Off-line reconstruction.

Issue 10: Issue with the GeneratorMonitor state machine in RHost code -

When the operator turns the key switch on the CT box to ‘Off’ and turns it back on before the system gantry is turned off completely the system might make the generator to go into a configuration state and ceases to monitor and to send the updated real time data for the tube’s kV and mA.

Affected products

  1. BRILLIANCE 40/40 SLICE CT SYSTEM

Lot or serial number

Not Applicable

Model or catalog number

728231

Companies
Manufacturer

Philips Medical Systems (Cleveland) Inc.

595 MINER ROAD

CLEVELAND

44143

Ohio

UNITED STATES



B. INGENUITY CT SYSTEM

Lot or serial number

Not Applicable

Model or catalog number

728323

728326

Companies
Manufacturer

Philips Medical Systems (Cleveland) Inc.

595 MINER ROAD

CLEVELAND

44143

Ohio

UNITED STATES