Brilliance 40-64 Slice CT System, Brilliance ICT System - System, Brilliance ICT SP System - Main, Ingenuity CT System (2019-06-25)
- Starting date:
- June 25, 2019
- Posting date:
- July 19, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-70453
Last updated:
Affected Products
A. Brilliance 40-64 Slice CT System
B. Brilliance ICT System- System
C. Brilliance ICT SP System - Main
D. Ingenuity CT System
Reason
There are 3 scenarios users may experience which result in identical images that are incorrectly labeled as phase tolerance images:
- When performing a Step and Shoot acquisition with phase tolerance selected, images may be reconstructed identically (a single phase), but are labeled as different phases. This occurs when ""Start Final Recon"" is selected prior to display of the ECG wave at the top of the acquisition window.
- When performing a Step and Shoot acquisition with phase tolerance selected, images may be reconstructed identically (a single phase), but are labeled as different phases. This occurs if ECG leads become disconnected mid-acquisition or when the acquisition is halted prematurely due to an application crash.
- By system design, in certain instances when a patient exhibits an arrhythmia or varying heart rate during a Step and Shoot acquisition, images from the acquisition following the heart rate variation, may be labeled in the incorrect phase.
Affected products
A. Brilliance 40-64 Slice CT System
Lot or serial number
Not applicable.
Model or catalog number
728231
Companies
- Manufacturer
-
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
B. Brilliance ICT System- System
Lot or serial number
Not applicable.
Model or catalog number
728306
Companies
- Manufacturer
-
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
C. Brilliance ICT SP System - Main
Lot or serial number
Not applicable.
Model or catalog number
728311
Companies
- Manufacturer
-
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
D. Ingenuity CT System
Lot or serial number
Not applicable.
Model or catalog number
728323
728326
Companies
- Manufacturer
-
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS