Health product recall

Bridge Occlusion Balloon (2020-03-30)



Starting date:
March 30, 2020
Posting date:
April 17, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-72815


Last updated:

Affected Products

Bridge Occlusion Balloon

Reason

There is a potential for thrombus formation on balloons prepped and staged in the vasculature for long durations (mean 176±118Min).   The formation of thrombus by foreign objects placed in the vasculature is a known phenomenon and would be expected to occur with any balloon left dwelling in a vein for an extended period of time.  While this is common knowledge for physicians that routinely do catheter based work (i.e., IRS, ICS), we have come to realize this is not common knowledge for the physicians (i.e., electrophysiologists) performing lead extraction cases and using these balloons as rescue devices.

Affected products

Bridge Occlusion Balloon

Lot or serial number

Not applicable.

Model or catalog number

590-001

Companies
Manufacturer

Spectranetics Corporation

9965 FEDERAL DRIVE

COLORADO SPRINGS

80921

Colorado

UNITED STATES