BP-200 Plus Blood Pressure Monitor
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
BP-200 Plus Blood Pressure Monitor |
More than 10 numbers, contact manufacturer. |
5.310011 |
Issue
Masimo identified TC- I reusable tip clip sensors with higher-than-expected variability in SpM2 readings that are outside of the accuracy specification. Masimo's investigation found the TC-I sensors had higher variability with an overall arms of 4.4% compared to the specification of 3.5% arms. These sensors are included with the BP-200 plus device with the SpO2 option.
Recall start date: March 14, 2024