Health product recall

Bodyguard pumps (2021-09-16)



Starting date:
September 16, 2021
Posting date:
October 8, 2021
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-76601


Last updated:

Affected Products

A. Bodyguard Twins - Pump

B. Bodyguard 323 IV Infusion Pump | Bodyguard 323 -510 refurbished Pump kit

Reason

Based on customer feedback, CME has identified potential flow rate issues when using the CME bodyguard microsets/bodysets infusion sets, with the associated CME bodyguard infusion pump. The investigation showed an increased risk of under-infusion when using the pump system at high flow rate settings in both continuous and an intermittent infusion like PCA mode. Deviations of nominal accuracy (5-7%) were detected above 500 Ml/Hr and were most prevalent and significant (20%) When running infusions at flow rates 800 Ml/hr in continuous mode. Similarly, when PCA Bolus (intermittent) are delivered at rates 500 Ml/hr in combination with low basal rates. The investigation showed an increased risk of under-infusion when using the pump system at high flow rate settings in both continuous and an intermittent infusion like pca mode. Deviations of nominal accuracy (5-7%) were detected above 500 ml/Hr and were most prevalent and significant (20%) When running infusions at flow rates 800 Ml/Hr in continuous mode. Similarly, when PCA Bolus (intermittent) are delivered at rates  500 Ml/Hr in combination with low basal rates.

Affected products

A. Bodyguard Twins - Pump

Lot or serial number

More than 10 numbers, contact manufacturer. 

Model or catalog number

TWINS 121

Companies
Manufacturer

Caesarea Medical Electronics Ltd.

16 Shacham Street, Industrial Park P. O. Box 4294

Caesarea

38900

ISRAEL



B. Bodyguard 323 IV Infusion Pump | Bodyguard 323 -510 refurbished Pump kit

Lot or serial number

More than 10 numbers, contact manufacturer. 

Model or catalog number

BODYGUARD 323

Companies
Manufacturer

Caesarea Medical Electronics Ltd.

16 Shacham Street, Industrial Park P. O. Box 4294

Caesarea

38900

ISRAEL