This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Blood Grouping Reagent MTS A/B/D Monoclonal and Reverse Grouping Card (2014-07-07)

Starting date:
July 7, 2014
Posting date:
July 25, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-40629

Recalled Products

Blood Grouping Reagent MTS A/B/D Monoclonal and Reverse Grouping Card

Reason

Ortho Clinical Diagnostics (OCD) observed an increased number of complaints related to false positive and/or reactions containing haze in the Anti-A and buffered gel microtubes used for reverse ABO typing in the MTS A/B/D Monoclonal and Reverse Grouping Card lots listed above. Our investigation determined that the false positive reactions were due to broken gel beads in the microtubes of the MTS cards.

Affected products

Blood Grouping Reagent MTS A/B/D Monoclonal and Reverse Grouping Card

Lot or serial number

02413037-53

120213037-55

Model or catalog number

MTS080515

Companies
Manufacturer

Micro Typing Systems Inc.

1295 29th Ave. S.W.

Pompano Beach

33069

Florida

UNITED STATES