This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

BIRMINGHAM HIP RESURFACING SYSTEM (2015-06-04)

Starting date:
June 4, 2015
Posting date:
June 26, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Identification number:
RA-53952

Recalled Product

A.BIRMINGHAM HIP RESURFACING SYSTEM (BHR) -  ACETABULAR CUPS

B.BIRMINGHAM HIP RESURFACING SYSTEM (BHR) -  DYSPLASIA CUPS

C.BIRMINGHAM HIP RESURFACING SYSTEM (BHR) -  FEMORAL HEADS

Reason

As part of its post market surveillance (PMS) and post-marketing clinical follow-up processes, Smith & Nephew has conducted an analysis of recent National Joint Registry of England and Wales (NJREW) data, (the largest arthroplasty registry cohort of BHR patients). A health hazard evaluation (HHE) was conducted to review this analysis. The data indicate that the BHR System continues to perform well in the male population requiring femoral head components 50mm in diameter and larger. However, the revision rates associated with the female gender, and smaller femoral head sizes regardless of gender, perform less well and exceed the current revision rate benchmark established by the UK National Institute for Health and Care Excellence (NICE).

Affected products

A.BIRMINGHAM HIP RESURFACING SYSTEM (BHR) -  ACETABULAR CUPS

Lot or serial number

All

Model or catalog number

74120144

74120146

74120148

74120150

74120152

74120154

74122146

74122148

74122150

74122152

Companies
Manufacturer

Smith & Nephew Orthopaedics Ltd.,

Harrison Way,

Leamington Spa, Warwickshire

CV31 3HL

UNITED KINGDOM



B.BIRMINGHAM HIP RESURFACING SYSTEM (BHR) -  DYSPLASIA CUPS

Lot or serial number

All

Model or catalog number

74120246                          74120250                            74120254                            74122248                           74122252

Companies
Manufacturer

Smith & Nephew Orthopaedics Ltd.,

Harrison Way,

Leamington Spa, Warwickshire

CV31 3HL

UNITED KINGDOM



C.BIRMINGHAM HIP RESURFACING SYSTEM (BHR) -  FEMORAL HEADS

 

Lot or serial number

All

Model or catalog number

74121138

74121142

74121146

74123140

74123144

 

Companies
Manufacturer

Smith & Nephew Orthopaedics Ltd.,

Harrison Way,

Leamington Spa, Warwickshire

CV31 3HL

UNITED KINGDOM