Health product recall

BIORAPTOR™ Suture Anchors

Last updated

Summary

Product
BIORAPTOR™ Suture Anchors
Issue
Medical devices - Sterility
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Bioraptor Curved 2.3 Pk Suture Anchor

2151692
2151422

72203281
72203280

Bioraptor 2.3 Pk Suture Anchor W/Ultrabraid Suture

2151690
2151242
2146369
2151691
2153489

72201542
72201541

Issue

Smith & Nephew, Inc. has initiated a field action to voluntarily remove seven (7) lots of BIORAPTOR™ Suture Anchors due to a potential for sterile barrier breach. Under certain shipping conditions, the distal end of the inserter can potentially pierce the packaging tray, compromising the sterile barrier. 

Recall start date: October 29, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Orthopaedics
Companies
Smith & Nephew, Inc. Endoscopy
150 Minuteman Road, Andover, Massachusetts, United States, 01810
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76459

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe