Health product recall

Biograph mCT-S with TrueV (64) and TrueV (40), Biograph mCT-S (40) and (20), Biograph mCT Flow-3R, Flow-4R and Flow Edge-4R (2018-05-16)

Starting date:
May 16, 2018
Posting date:
June 22, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-67072

Affected products

A. Biograph mCT-S with TrueV (64)
B. Biograph mCT-S (40)
C. Biograph mCT-S with TrueV (40)
D. Biograph mCT Flow-3R
E. Biograph mCT Flow-4R
F. Biograph mCT Flow Edge-4R
G. Biograph mCT-S (20)

Reason

Siemens is notifying users of Biograph Systems of an issue that could potentially impact the CARE Dose4D algorithm in CT scans that include the head. An increase in dose is possible when scanning the head region with CARE Dose4D using a Posterior-Anterior (PA) or Anterior Posterior (AP) topogram. The customer letter includes instructions to avoid the issue.

Affected products

A. Biograph mCT-S with TrueV (64)

Lot or serial number

11030

Model or catalog number

10248672

Companies
Manufacturer
Siemens Medical Solutions USA, Inc. Molecular Imaging
2501 North Barrington Road
Hoffman Estates
60192
Illinois
UNITED STATES

B. Biograph mCT-S (40)

Lot or serial number

11044

Model or catalog number

10248668

Companies
Manufacturer
Siemens Medical Solutions USA, Inc. Molecular Imaging
2501 North Barrington Road
Hoffman Estates
60192
Illinois
UNITED STATES

C. Biograph mCT-S with TrueV (40)

Lot or serial number
  • 11058
  • 11069
Model or catalog number

10248671

Companies
Manufacturer
Siemens Medical Solutions USA, Inc. Molecular Imaging
2501 North Barrington Road
Hoffman Estates
60192
Illinois
UNITED STATES

D. Biograph mCT Flow-3R

Lot or serial number

59002

Model or catalog number

10528956

Companies
Manufacturer
Siemens Medical Solutions USA, Inc. Molecular Imaging
2501 North Barrington Road
Hoffman Estates
60192
Illinois
UNITED STATES

E. Biograph mCT Flow-4R

Lot or serial number
  • 11002
  • 11010
  • 60036
Model or catalog number
  • 10529159
  • 10529161
Companies
Manufacturer
Siemens Medical Solutions USA, Inc. Molecular Imaging
2501 North Barrington Road
Hoffman Estates
60192
Illinois
UNITED STATES

F. Biograph mCT Flow Edge-4R

Lot or serial number

11013

Model or catalog number

10528955

Companies
Manufacturer
Siemens Medical Solutions USA, Inc. Molecular Imaging
2501 North Barrington Road
Hoffman Estates
60192
Illinois
UNITED STATES

G. Biograph mCT-S (20)

Lot or serial number

21147

Model or catalog number

10507786

Companies
Manufacturer
Siemens Medical Solutions USA, Inc. Molecular Imaging
2501 North Barrington Road
Hoffman Estates
60192
Illinois
UNITED STATES