Health product recall

BIOGRAPH MCT and BIOGRAPH HORIZON System (2018-10-19)

Starting date:
October 9, 2019
Posting date:
October 25, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-71385


Last updated:

Affected Products

  1. BIOGRAPH MCT-S WITH TRUE V (64)
  2. BIOGRAPH MCT-S (40)
  3. BIOGRAPH MCT-S WITH TRUE V (40)
  4. BIOGRAPH MCT FLOW-3R
  5. BIOGRAPH MCT FLOW-4R
  6. BIOGRAPH MCT FLOW EDGE-4R
  7. BIOGRAPH MCT-S (20)

Reason

Siemens Medical Solutions USA Inc., Molecular imaging determined there is potential that the screws to the system power connection terminal may be loose. The issue affects BIOGRAPH MCT and BIOGRAPH HORIZON systems. Siemens Healthineers is initiating a field action to inspect potentially affected systems and ensure the power connector screws are torqued to the proper specification.

Affected products

A. BIOGRAPH MCT-S WITH TRUE V (64)

Lot or serial number

11030

Model or catalog number

10248672

Companies
Manufacturer

Siemens Medical Solutions USA, Inc. Molecular Imaging

2501 North Barrington Road

Hoffman Estates

60192

Illinois

UNITED STATES



B. BIOGRAPH MCT-S (40)

Lot or serial number

11044

Model or catalog number

10248668

Companies
Manufacturer

Siemens Medical Solutions USA, Inc. Molecular Imaging

2501 North Barrington Road

Hoffman Estates

60192

Illinois

UNITED STATES



C. BIOGRAPH MCT-S WITH TRUE V (40)

Lot or serial number
  • 11058
  • 11069
Model or catalog number

10248671

Companies
Manufacturer

Siemens Medical Solutions USA, Inc. Molecular Imaging

2501 North Barrington Road

Hoffman Estates

60192

Illinois

UNITED STATES



D. BIOGRAPH MCT FLOW-3R

Lot or serial number

59002

Model or catalog number

10528956

Companies
Manufacturer

Siemens Medical Solutions USA, Inc. Molecular Imaging

2501 North Barrington Road

Hoffman Estates

60192

Illinois

UNITED STATES



E. BIOGRAPH MCT FLOW-4R

Lot or serial number
  • 11010
  • 60036
Model or catalog number

10529161

Companies
Manufacturer

Siemens Medical Solutions USA, Inc. Molecular Imaging

2501 North Barrington Road

Hoffman Estates

60192

Illinois

UNITED STATES



F. BIOGRAPH MCT FLOW EDGE-4R

Lot or serial number

11013

Model or catalog number

10528955

Companies
Manufacturer

Siemens Medical Solutions USA, Inc. Molecular Imaging

2501 North Barrington Road

Hoffman Estates

60192

Illinois

UNITED STATES



G. BIOGRAPH MCT-S (20)

Lot or serial number
  • 11002
  • 21147
Model or catalog number
  • 10507786
  • 10529159
Companies
Manufacturer

Siemens Medical Solutions USA, Inc. Molecular Imaging

2501 North Barrington Road

Hoffman Estates

60192

Illinois

UNITED STATES