Health product recall

Biofire® Filmarray® Torch System Base

Last updated

Summary

Product
Biofire® Filmarray® Torch System Base
Issue
Medical devices - Performance
What to do

Contact the importer.

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Biofire® Filmarray® Torch System Base

All lots.

HTFA-ASY-0104

Issue

When turning on biofire torch Systems, arcing inside of the power switch may result in carbon build‐up on the switch contacts. The carbon build‐up may lead to excess heating inside of the power switch, subsequently leading to deformation of the power switch case. The deformation can create an opportunity for a power switch electrical short. In addition to an electrical short, deformation may result in an open circuit causing the power switch to fail. This event would only occur after the product is in use.

Recall Start Date: October 27, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Companies

Biofire Diagnostics LLC

515 Colorow Dr., Salt Lake City, Utah, United States, 84108

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-74575