This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

BioArray Solution's RHD BeadChip Kit (2016-02-11)

Starting date:
February 11, 2016
Posting date:
March 4, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-57420

Affected Products

BioArray Solution's RHD BeadChip Kit

Reason

BioArray Solutions Ltd. Has determined that RHD master lot 16-166 has incorrect error cutoff values. Lot-specific error cutoff values for the RHD SNP markers released into the basis global software for RHD master lot 16-166 can create indeterminate genotype calls ("XX") which may impact the accuracy of the variant allele calls.

Affected products

BioArray Solution's RHD BeadChip Kit

Lot or serial number

16-166

Model or catalog number

800-10220-48

Companies
Manufacturer

BIOARRAY SOLUTIONS LTD

35 TECHNOLOGY DRIVE, SUITE 100,

Warren

07059

New Jersey

UNITED STATES