Berhring Nephelometer SystemN Latex CDT Kit (Carb Deficient Transferrin) (2020-03-05)
- Starting date:
- March 5, 2020
- Posting date:
- April 15, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-72715
Last updated:
Affected Products
Berhring Nephelometer SystemN Latex CDT Kit (Carb Deficient Transferrin)
Reason
Siemens Healthcare Diagnostics has observed a negative bias for the carbohydrate-deficient transferrin (CDT) measurement when using the affected N latex CDT kit lots (49729, 49849, 49985 and 50084) in comparison to the HPLC method. All affected kit lots contain reagents from the same bulk. The issue has been internally observed during performed shelf-life monitoring after 15 months. CDT results in absolute concentrations may be influenced by the patient's transferrin levels and, therefore, results are reported as a ratio of CDT to total transferrin, called
Affected products
Berhring Nephelometer SystemN Latex CDT Kit (Carb Deficient Transferrin)
Lot or serial number
- 49729
- 50084
Model or catalog number
OPCS03