Benchmark Ultra Stainer Module
Brand(s)
Last updated
Summary
Product
Benchmark Ultra Stainer Module
Issue
Medical devices - Manufacturing defect
What to do
Contact the manufacturer if you require additional information.
Affected products
Affected Products |
Lot or Serial Number |
Model or catalog number |
|---|---|---|
Benchmark Ultra Stainer Module |
More than 10 numbers, contact manufacturer. |
N750-BMKU-FS |
Issue
The manufacturer reported a fluid leak within the Benchmark Ultra instruments that created an electrical short on the instrument's general purpose input/output (GPIO) board's printed circuit board (PCB). The event was localized to the J1 connector of the GPIO board in the lower rear portion inside the instrument and there was no injury.
Recall start date: Jan 18, 2022
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Pathology
Companies
1910 E Innovation Park Drive, Tucson, Arizona
United States, 85755
Published by
Health Canada
Audience
General public
Health professionals
Hospitals
Recall class
Type II
Identification number
RA-63904