Health product recall

Bekman Coulter AU Immunochemistry System-AU480 (2013-09-03)

Starting date:
September 3, 2013
Posting date:
September 16, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-41383

Recalled products

  1. Bekman Coulter AU Immunochemistry System-AU480

Reason

When the AU480 Clinical Chemistry Analyzer operates using the real-time online function, and the number of barcode digits in the sample ID barcode label does not match the instrument configuration setting in the "requisition format" menu and in the LIS system, the AU480  Clinical Chemistry Analyzer generates alarm 6042 "online mismatch" and the sample is not processed. If the number of barcode digits configured in the LIS, middle ware, and AU480 settings match, the customer will not experience the issue. The issue does not generate incorrect results but may cause a delay in processing the sample.

Affected products

A. Bekman Coulter AU Immunochemistry System-AU480

Lot or serial number
  • 2013072465
  • 2013082468
Model or catalog number
  • AU480
Companies
Manufacturer

Beckman Coulter Inc.

250 S. Kraemer Blvd.

Brea

92821

California

UNITED STATES