Bekman Coulter AU Immunochemistry System-AU480 (2013-09-03)
- Starting date:
- September 3, 2013
- Posting date:
- September 16, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-41383
Recalled products
- Bekman Coulter AU Immunochemistry System-AU480
Reason
When the AU480 Clinical Chemistry Analyzer operates using the real-time online function, and the number of barcode digits in the sample ID barcode label does not match the instrument configuration setting in the "requisition format" menu and in the LIS system, the AU480 Clinical Chemistry Analyzer generates alarm 6042 "online mismatch" and the sample is not processed. If the number of barcode digits configured in the LIS, middle ware, and AU480 settings match, the customer will not experience the issue. The issue does not generate incorrect results but may cause a delay in processing the sample.
Affected products
A. Bekman Coulter AU Immunochemistry System-AU480
Lot or serial number
- 2013072465
- 2013082468
Model or catalog number
- AU480
Companies
- Manufacturer
-
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES