Becton Dickinson Vacutainer Plus Plastic Plasma Separator Tube (PST)
- Starting date:
- April 2, 2013
- Posting date:
- June 17, 2013
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-34175
Recalled products
- Becton Dickinson Vacutainer Plus Plastic Plasma Separator Tube (PST)
Reason
A product defect named "lip-out" was detected for BD Vacutainer PST tubes. It consists of gaps of various sizes in the structure of the tube wall that is present at the lip of the tube. These gaps are covered by the stopper and hemogard closure, so the tube can be evacuated and used for blood collection. Customer complaints cite jamming of preanalytical automation (tube decappers) and leakage of blood in the specimen transport bag.
Affected products
A. Becton Dickinson Vacutainer Plus Plastic Plasma Separator Tube (PST)
Lot or serial number
More than 10 numbers, contact manufacturer
Model or catalog number
- 367960
- 367961
- 367962
Companies
- Manufacturer
-
Becton Dickinson and Company
1 Becton Drive
Franklin Lakes
07417
New Jersey
UNITED STATES