This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Becton Dickinson Vacutainer Plus Plastic Plasma Separator Tube (PST)

Starting date:
April 2, 2013
Posting date:
June 17, 2013
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-34175

Recalled products

  1. Becton Dickinson Vacutainer Plus Plastic Plasma Separator Tube (PST)

Reason

A product defect named "lip-out" was detected for BD Vacutainer PST tubes. It consists of gaps of various sizes in the structure of the tube wall that is present at the lip of the tube. These gaps are covered by the stopper and hemogard closure, so the tube can be evacuated and used for blood collection. Customer complaints cite jamming of preanalytical automation (tube decappers) and leakage of blood in the specimen transport bag.

Affected products

A. Becton Dickinson Vacutainer Plus Plastic Plasma Separator Tube (PST)

Lot or serial number

More than 10 numbers, contact manufacturer 

Model or catalog number
  • 367960
  • 367961
  • 367962
Companies
Manufacturer

Becton Dickinson and Company

1 Becton Drive

Franklin Lakes

07417

New Jersey

UNITED STATES