Health product recall

Beckman Coulter Au Immunochemistry System - Bicarbonate (CO2) Assay

Last updated

Summary

Product
Beckman Coulter Au Immunochemistry System - Bicarbonate (CO2) Assay
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Beckman Coulter Au Immunochemistry System - Bicarbonate (CO2) Assay

All lots.

OSR6237
OSR6137

Beckman Coulter Au2700 Immunochemistry System - Bicarbonate(CO2) Assay

All lots.

OSR6237
OSR6137

Beckman Coulter Au5400 Immunochemistry System - Bicarbonate (CO2)Assay

All lots.

OSR6237
OSR6137

Issue

Beckman Coulter has confirmed a premature decrease in calibration ODs for several lots of bicarbonate OSR6X37. It is possible that all bicarbonate lots may also be impacted by this issue, therefore all lots are subject to this field action.

The abnormal decrease in calibrator ODs may result in calibration failures and/or OC failures. the risk of generating falsely elevated, clinically impactful patient test results is remote.

Recall start date: April 3, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Companies

Beckman Coulter, Inc.

250 S. Kraemer Blvd., Brea, California, United States, 92821

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-73500

Get notified