Beckman Coulter Au Immunochemi Sys- Unsaturated Iron Binding Capacity (UIBC)
Brand(s)
Last updated
Summary
Product
Beckman Coulter Au Immunochemi Sys- Unsaturated Iron Binding Capacity (UIBC)
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
Beckman Coulter Au Immunochemi Sys- Unsaturated Iron Binding Capacity (UIBC) |
All lots. |
OSR61205 |
Issue
UIBC OSR61205 is not meeting interference claims for haemoglobin interference serum application. As per product ifus the claim is less than 10% up to 2g/l haemoglobin. While the high pool is meeting the claim, the low pool is failing the hemolysate interference specifications with results of -43.63% or -8.95 μmol/l.
Recall start date: July 1, 2025
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Companies
Beckman Coulter, Inc.
250 S. Kraemer Blvd., Brea, California, United States, 92821
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-77710
Get notified
Receive emails about new and updated recall and safety alerts.