Health product recall

BD Trucount tubes (2016-06-27)

Starting date:
June 27, 2016
Posting date:
July 22, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
Healthcare Professionals, General Public, Hospitals
Identification number:
RA-59472

Affected products

A. BD Trucount tubes

Reason

Pouches contained in the kit lot 6083708 may not be sealed properly. The pouches contained in the kit can be identified with the lot no.16040 and are embossed with "17 02 16 t3" or "16 02 16 T3". Improper storage of Trucount Tubes due to failed pouch closure can affect the performance of the counting beads leading to erroneous results in enumeration assays utilizing Trucount. If the pouch seal is intact and the desiccant included in the pouch is blue, the product may be used. If the pouch seal integrity is compromised and/or the desiccant has turned from blue to lavender, discard the tubes, because exposure to the ambient environment can rapidly degrade the performance of the product.

Affected products

A. BD Trucount tubes

Lot or serial number

6083708

Model or catalog number

340334

Companies
Manufacturer

Becton, Dickinson and Company

BD Biosciences

2350 Qume Drive

San Jose

95131

California

UNITED STATES