This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

BD PCR Cartridges 24

Last updated

Summary

Product
BD PCR Cartridges 24
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare
Industry

Affected products

Affected products

Lot or serial number

Model or catalogue number

BD PCR Cartridges 24

More than 10 numbers, contact manufacturer.

437519

Issue

BD has identified through customer complaint investigations, that signal drift has been observed in specific lots of BD PCR cartridges. BD has identified the following impacts:
• for the BD Max™ RVP assay (catalog # 445215 & 445373) signal drift could cause false positive results for Influenza A.
• for specific BD Max™ molecular assays including enteric bacterial panel (EBP, catalog #442963), enteric parasite panel (EPP, catalog #442960), StaphSR (catalog
#443419), MDR-TB (catalog #443878), CPO (catalog # 445262) signal drift could potentially cause an increase in unresolved results (UNR).
• for the BD COR™ VP (catalog # 443976) signal drift could potentially cause an increase in unresolved results (UNR).
• for third-party assays used with the open system reagent kits (BD MAX™ ExK™ DNA-
1, DNA-2, DNA-3, DNA MMK, DNA MMK (SPC), TNA MMK, TNA MMK (sSPCpc), TNA-2 and TNA-3) signal drift could potentially cause an increase in unresolved results (UNR) or unexpected results.

Recall start date: December 19, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Microbiology
Companies
Becton Dickinson And Company (Bd)
7 Loveton Circle, Sparks, Maryland, United States, 21152
Published by
Health Canada
Audience
Healthcare
Industry
Recall class
Type II
Identification number
RA-76789

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe