This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

BC-THROMBIN REAGENT (2017-10-12)

Starting date:
October 12, 2017
Posting date:
October 27, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
Healthcare Professionals, Hospitals, General Public
Identification number:
RA-64962

Affected products

BC-THROMBIN REAGENT

Reason

The BC Thrombin Reagent Kit lots 46751 (contains Thrombin Reagent lot 517468) and Kit lot 47184 (contains Thrombin Reagent lot 517469) produce unexpected prolonged thrombin time results for expected normal samples, and may recover above the upper limit of normal which is 21 sec as stated within the IFU (instruction for use) and the BCS/ BCS XP System application sheets.

Affected products

BC-THROMBIN REAGENT

Lot or serial number
  • 46751
  • 47184
Model or catalog number

OWNA11

Companies
Manufacturer

Siemens Healthcare Diagnostics Products Gmbh

Emil-Von-Behring-Str. 76

Marburg

35041

GERMANY